Perma Pure MiniGASS
Request a quote
Add this product to the quotation list or request a quotation for this product directly. Within two business days you'll receive a detailed quotation drafted by one of our experts.
Product Description
MiniGASS™ sample conditioning systems combine coalescing filtration,
PID-controlled heating and Nafion™ gas-phase drying down to -25°C in one compact
unit. Suitable for CEMS installations, process analysis and OEM integration where
reliable sample preparation is essential.
Product specifications
| Product specifications | Test method | ||
|---|---|---|---|
| Application | Measuring and controlling gases & liquids | ||
| Product group | Gas dryers (Nafion) | ||
Perma Pure MiniGASS™ Gas Sampling Systems
Integrated 3-in-1 sample conditioning for CEMS and industrial gas analysis
MiniGASS™ sample conditioning systems combine coalescing filtration, PID-controlled heating and Nafion™ gas-phase drying down to -25°C in one compact unit. This integrated approach eliminates condensation, protects analysers against contamination and retains water-soluble analytes, which is essential for reliable CEMS measurements, process analysis and OEM integration.
The MiniGASS series prepares hot, humid and contaminated gas samples by removing particles, acid mists and water vapour step by step without analyte loss. This results in reduced analyser downtime, extended service intervals and lower maintenance costs in challenging industrial environments.
Technical operating principle
MiniGASS systems condition gas samples in an integrated three-step process:
1. Sample filtration Coalescing filter elements effectively remove particles, mist and aerosols from the gas stream. This prevents clogging, corrosion and damage to downstream analysers. Filter elements are available in different pore sizes (0.01-20 µm) and materials for optimal performance with specific gas compositions.
2. Heating and temperature control An integrated PID temperature controller with display controls the heated filter/probe assembly (480W silicone pad heater). This keeps the sample above the dew point during pre-conditioning and prevents condensation, even at inlet temperatures up to 275°F (135°C).
3. Nafion™ gas-phase drying Nafion™ membrane technology selectively removes water vapour from the gas stream down to dew points of -25°C or lower. This gas-phase drying retains water-soluble analytes (SO₂, NH₃, HCl, HF) and prevents the analyte loss that occurs with condensation-based methods. Available in three capacities: 50-tube (≤1 LPM), 100-tube (2-3 LPM) or 200-tube (4-6 LPM).
Technical advantages
Integrated 3-in-1 solution
- Complete sample conditioning in one compact unit reduces installation complexity and commissioning time
- Eliminates the need for separate filtration, heating and drying systems
- Standardised interface simplifies OEM integration
Drying to very low dew points
- Nafion™ drying down to -25°C prevents condensation in sample lines and analyser optical cells
- Eliminates water-vapour-related spectral interference in FTIR and NDIR analysers
- Reduces CEMS downtime through stable, dry gas samples
Retention of water-soluble analytes
- Gas-phase drying without condensation retains SO₂, NH₃, HCl, HF and other soluble components
- Essential for accurate emission measurements according to IED and BREF guidelines
- Prevents analyte loss and measurement inaccuracies
Extended analyser service intervals
- Effective particle and mist removal (down to 0.01 µm) protects analyser components
- Typical extension of maintenance intervals: 3-12 months depending on gas composition
- Reduced costs through fewer analyser failures and lower spare parts consumption
Chemical resistance for corrosive applications
- NEMA Type 4X enclosure for aggressive environments (-20°C to 40°C ambient)
- Fluorocarbon wetted materials (PVDF/FEP) or 316SS for SO₂, NOₓ, HCl, HF and acid gas streams
- FKM O-rings and glass-fibre coalescing filters for maximum service life
Technical specifications
Physical
| Parameter | Specification |
|---|---|
| System mount | Wall mounted |
| Enclosure ratings | UL-508A Listed; Type 4, 4X, 12, 13; File No. E61997cUL Listed per CSA C22.2 No 94; Type 4, 4X, 12, 13; File No. E61997 |
| Sample gas tubing connections | 1/4" or 3/8" tube compression, Kynar® (PVDF) or 316SS |
| Umbilical line seal (optional) | Heat shrink style, 2" or 3" min expanded I.D. |
| Weight | Approximately 45lb (24.4kg) depending on selected options |
Utility requirements
| Parameter | Specification |
|---|---|
| Electrical requirement | 115V ±10% / 230V ±10%, 5.0A/2.5A, 50/60Hz |
| Compressed/instrument air requirements | 30-100 psig (2-7 bar), 1 scfm (28 slpm) instrument air vereist alleen wanneer optionele "Heatless air dryer" is opgenomen |
| 60-100 psig (4-7 bar), 2 scfm (56 slpm) perslucht vereist wanneer optionele "Heatless dryer" is opgenomen | |
| Aanvullende 2 scfm (56 slpm) vereist als koellucht wanneer sample pump is opgenomen | |
| Drain/exhaust requirement | 25 scfm (7 slpm) air with entrained acidic mist. Works in a timed cycle. Not continuous. |
Environmental
| Parameter | Specification |
|---|---|
| Temperature range, operational | -20°C to 40°C ambient temperature |
| Humidity range, operational | 0-95% RH |
Sample gas limits
| Parameter | Specification |
|---|---|
| Temperature | 275°F (135°C) |
| Pressure | Maximum 20 psig |
| Minimum 5" H2O vacuum without purge eductor (9-NE option) | |
| Minimum 5" Hg vacuum with purge eductor (standard) | |
| Humidity | Maximum 30% by vol., <63C dew point |
Control and Electrical
| Parameter | Specification |
|---|---|
| Temperature controller | PID type with low temperature alarm |
| Heater type | 480watt silicone pad heater on enclosure back panel |
| Alarm outputs | SPDT relays, 3A, 250VAC (de-energised in alarm state). Screw terminal connection |
| One set of SPDT relay contacts available for low purge air flow alarm | |
| One set of SPDT relay contacts available for low temperature alarm |
Materials
| Component | Material |
|---|---|
| System enclosure | Glass-fibre reinforced polyester |
| Insulation | Silicone rubber foam |
| O-ring seals | FKM |
| Sample gas path fittings and tubing | PVDF/FEP or 316SS |
| Filter housing | PVDF (PVDF) / glass shell or 316 SS / glass shell |
| Filter holder, coalescing filter element | Glass fibre, 0.01 µm, 95% efficiency, coalescing |
System configurations
MiniGASS 1228W: wall mount system
Wall-mount version in a robust NEMA Type 4X glass-fibre reinforced polyester enclosure with integrated coalescing filter, PID temperature control, purge gas flowmeter, pressure regulator and digital display. Suitable for CEMS analyser shelters, process analysis cabinets and outdoor installations. UL-508A Listed and CSA C22.2 certified.
MiniGASS 1228P: probe mount system
Probe-mount version for direct stack installation with a transparent enclosure for visual inspection during operation. Complete instrumentation including purge gas flowmeter and PID controller. Compact footprint for mounting in confined spaces with flange options from 2" to 6" (150#). Optional gas cooling spools available for high process temperatures.
Fields of application
Emission monitoring (CEMS) - compliance & reporting
Energy sector - coal, gas and biomass power plants. Continuous SO₂, NOₓ, CO, CO₂, O₂ measurements according to the IED (Industrial Emissions Directive), BREF documents and national guidelines for emission reporting.
Waste processing - waste incineration plants, medical waste incinerators and hazardous waste incineration. Compliance monitoring for HCl, HF, heavy metals and dioxin-related parameters.
Industry - cement kilns, steel plants, glass furnaces and the ceramics industry. Process emission monitoring and EU ETS reporting for CO₂ emissions trading.
Process analysis & optimisation
Chemical industry - reactor off-gas monitoring, process optimisation and safety monitoring of toxic components. Real-time feedback for process control in high-humidity environments.
Petrochemicals - refinery process streams, flare gas monitoring and catalytic cracker units. Sample conditioning for corrosive gas streams with H₂S, SO₂ and other acid-forming components.
OEM integration & stack testing
FTIR, NDIR, UV/VIS gas analysers, multi-component CEMS systems and portable emission monitoring systems (PEMS). Modular configuration for fast integration into analyser platforms and mobile stack testing equipment.
Ordering & technical support
For configuration advice, sizing and project support, please contact Inacom. Specify in your request:
- Gas composition and concentrations (including corrosive components)
- Sample flow rate, pressure and temperature
- Mounting environment (indoor/outdoor, hazardous area classification)
- Required dew point and analytical requirements (analyser type, measuring range)
- CEMS compliance requirements (IED, BREF, national guidelines)