Industrial oxygen and nitrogen analysers
Oxygen sensors in medical equipment have a limited service life and must be replaced periodically to guarantee reliable measurement. Maxtec supplies replacement sensors that are compatible with a wide range of existing equipment, from anaesthesia machines and ventilators to portable oxygen analysers and industrial gas monitors.
Max-11 The Max-11 is a galvanic replacement sensor compatible with anaesthesia and ventilation equipment from Dräger (Narkomed, Fabius GS, Evita, Babylog), EnviteC, Invivo, Teledyne, Teleflex and VTI, among others. Gold-plated slip-ring connection.
Measuring range: 0 to 100%, output signal 14 to 20 mV, response time ≤12 s, expected service life >500,000 %O₂ hours. Item number: R113P10, 12-month warranty
Max-13 The Max-13 is a galvanic replacement sensor with NTC temperature compensation, compatible with equipment from Covidien (7820 Monitor, 7200 Vent), Philips/Respironics (Criterion Vent, Esprit Vent), MSA (MiniOX 3000, MiniOX II and III), Dräger (Anestar S), Datascope and many others. Connection via 3.5 mm mono jack.
Measuring range: 0 to 100%, output signal 13.0 to 16.5 mV, response time ≤12 s, expected service life >1,000,000 %O₂ hours. Item number: R115P10, 14-month warranty
Max-80i
The Max-80i is an industrial oxygen sensor for custom OEM applications with a 3-pin Molex connection. Suitable for applications with high CO₂ concentrations or other acidic gases, applications that require a short response time and custom electrical interfaces.
Measuring range: 0 to 100 vol%, output signal 5 to 15 mV, response time <5 s, expected service life >1,200,000 %O₂ hours.
Compatibiliteitstabel
| Sensor | Vervangt o.a. | Connection |
|---|
| Max-11 | Dräger 6803290/6850645, EnviteC OOM-201, Teledyne R23, Teleflex 5557/V-14 | Gold-plated slip ring |
| Max-13 | Covidien 4-020933-00, Philips 37-079, MSA 406931, Dräger Anestar S, VTI 100580 | 3,5 mm mono-jack |
| Max-80i | Custom OEM, high-CO₂ environments, short response time required | 3-pin Molex |