Home Catalogue Pressure and flow control Mass flow meters and controllers SmartTrak® Series
SmartTrak® Series
Request a quote
Add this product to the quotation list or request a quotation for this product directly. Within two business days you'll receive a detailed quotation drafted by one of our experts.
Product Description
The SmartTrak® series is based on capillary mass flow technology, in which the mass flow is determined directly via the temperature differences along a thin capillary tube. This ensures excellent measurement accuracy, stability and repeatability, even with corrosive or aggressive gases.
Product specifications
| Product specifications | Test method | ||
|---|---|---|---|
| Application | Measuring and controlling gases & liquids | ||
| Product group | Mass flow meters and controllers | ||
Vögtlin SmartTrak® Series: Mass Flow Meters & Controllers
The Vögtlin SmartTrak® series consists of high-grade capillary mass flow meters (MFM) and mass flow controllers (MFC) for gases, specially developed for applications requiring high accuracy, chemical resistance and process stability. This series offers a reliable solution for both industrial processes and analytical systems that require high-grade gas flow measurement and control.
Key properties
Capillaire mass flow technologie
Thermal measurement via a capillary tube delivers very accurate mass flow data and a stable measuring signal, insensitive to pressure and temperature fluctuations.
Hoge meetnauwkeurigheid
SmartTrak® meters and controllers offer an accuracy of up to ±0,5 % of full scale, ensuring reliable results in critical applications.
316L roestvast staal natte delen
All gas-carrying parts are made of 316L stainless steel, making the SmartTrak® series ideal for dry, clean and corrosive gases.
Breed toepassingsgebied
Thanks to the combination of accuracy and choice of materials, SmartTrak® instruments are suitable for laboratory environments, process monitoring and industrial applications using aggressive media.
Analogue and digital interfaces
SmartTrak® units can be fitted with classic analogue outputs, but are also compatible with digital communication standards for integration into automated systems.
Compact and robust design
The series is designed for reliable operation in everyday applications and for integration into OEM systems or test setups.
Applications
The SmartTrak® series is widely used in situations such as:
- Measuring and controlling gas flows in laboratory environments
- Process control of aggressive or corrosive gases
- Precision gas dosing and gas distribution
- OEM-integratie in analytische instrumentatie
- Industries with high demands on measurement accuracy and chemical resistance
Thanks to the capillary measuring principle and the high-grade stainless steel version, this series is ideally suited for applications that require exceptionally reliable measurement.
Technical specifications
Measuring ranges | |
|---|---|
| Air/full scale freely selectable | SmartTrak® 100L (Low Flow): 0-10 mln/min to 0-50 ln/minSmartTrak® 100M (Medium Flow): 0-20 to 0-200 ln/minSmartTrak® 100H (High Flow): 0-100 to 0-500 ln/minFlow ranges specified are for an equivalent flow of nitrogen at 1013.25 mbara(760 mm Hg) and 0°C (32°F)For other gases and reference conditions please contact the factory. |
Instrument versions | |
|---|---|
| “Standard”The economic solution | Accuracy: ± 1.0% of full scale*Turndown ratio: 1 : 50 (Controller), 1 : 100 (Meter) |
| “Hi-Performance”With high accuracy | Accuracy: ± 0.5% of full scale*Including 10 point certificateHi-performance is only possible for:Flow range: up to 50 slpm or nlpm (valid from 10 to 100% of the calibrated range)Gases: Air, Nitrogen, Helium or ArgonOther gases and flow: Contact the factoryOnly valid for units up to an operating pressure of 10 barg and over the temperature range10°C to 30°C in a horizontal position. The accuracy is only guaranteed under calibrated conditions and is only valid for a period of 6 months after the date of calibration. Valid for 10 to 100% of FS. |
*An additional error of ±0.25% may apply for analogue signals
Performance data | |
|---|---|
| Media (real gas calibration) | All gases and gas mixtures that are compatible with the selected material. There might be limitations to some corrosive gases |
| Response time | 2 seconds (typical) to within ± 2% of final value (includes settling time),faster or slower available upon request (controllers only) |
| Repeatability | ± 0.2% of full scale (according to SEMI standard E56-0309) |
| Long-term stability | Typical < 2% of measured value / year |
| Power supply | Mass Flow Meters:15-24 VDC ±10%, (230 mA, regulated)Mass Flow Controllers:C100L: 24 VDC ±10% (500 mA, regulated)C100M: 24 VDC ±10%, (800 mA, regulated)C100H: 24 VDC ±10%, (1260 mA, regulated)(ripple should not exceed 100 mV peak-to-peak) |
| Operating pressure | 0-35 barg (0 – 500 psig) |
| Temperature (environment/gas) | 0 – 50°C (32 – 122°F) |
| Pressure sensitivity | <0.15% / bar (<0.01% / psi) FS (typical N2, horizontal mount) |
| Temperature sensitivity | <0.05% FS measuring range type per 1°C (<0.025% per 1°F) |
| Warm-up time | < 15 min for full accuracy |
Materials | |
|---|---|
| Meter all wetted metal parts | 316SS (1.4404) |
| Controller body/sensor materials | 316SS (1.4404) |
| Controllers valve post materials | C100L/C101: 316SS/410SS/430SS (1.4404/1.4006/1.4016)C100M: 416SS/430SS (1.4005/1.4016)C100H: 316SS/416SS/430SS (1.4404/1.4005/1.4016) |
| Electronic housing | Powder coated steel |
| Seals | FKM (Viton®) Options: Neoprene |
| Valve seals | FKM (Viton®) Options: Neoprene, Kalrex® |
Integration & installation | |
|---|---|
| Output signals analogue | Linear 4–20 mA or 0-20 mA, 500 ohms maximum loop resistanceand one of the following (user selectable):Linear 0–5 VDC, 1000 ohms minimum load resistanceLinear 0-10 VDC, 1000 ohms minimum load resistanceLinear 1-5 VDC, 1000 ohms minimum load resistance |
| Setpoint signals analog | Analog (choice of one, user selectable):Linear 4–20 mA, 0-20 mA. 0–5 VDC, 0-10 VDC, 1-5 VDC |
| Output signals digital | RS232 Optional: RS485; Modbus RTU (Slave) |
| Process connection | Fittings included. Options: Swagelok® (Compression), VCR, VCO, NPT |
| Inlet section | None required |
| Electrical connection | D-15 High density plug (15 pins) female |
| Mounting orientation | All orientations are possible but must be defined at time of order |
Safety | |
|---|---|
| Burst pressure | 52 bar (240 psi) |
| Leak rate | < 1 x 10-9 mbar l/s He |
| Ingress protection class | IP20 |
| EMC | EN 61326-1 |
| ATEX certification | None |