Peak broadening in HPLC: causes, dead volume and the Van Deemter equation

Published on 6 August 2026 Reading time : 5 minutes

Why your peaks become broader and which measures actually help

Peak broadening is the phenomenon in which an initially narrow sample band becomes increasingly broad during the analysis. The result: lower peaks, poorer resolution, less sensitivity and less reliable quantification. Peak broadening is therefore the most direct threat to the quality of your chromatographic data.

The causes can be divided into two groups: processes within the column, described by the Van Deemter equation, and extra-column effects caused by dead volume in the system. This article covers both and shows which measures have the greatest effect in practice.

⏱ approx. 8 minutes reading time

What is dead volume and why does it cause peak broadening?

Dead volume, also called extra-column volume, is all the volume in the system outside the column in which the sample can spread out. Every millilitre of it contributes to peak broadening without contributing to the separation.

Dead volume is found in these places:

  • The sample loop of the injection valve
  • All capillaries and tubing between injector, column and detector
  • The flow cell of the detector
  • Couplings, fittings and unions where the connection does not fit exactly

The effect proceeds in four steps: the sample spreads out before or after the column, the peak becomes broader, the peak height decreases because the same amount of substance is distributed over a longer time, and the resolution between neighbouring peaks falls.

In short

More dead volume gives more peak broadening and therefore poorer separation. With narrow columns and UHPLC applications, this effect is dramatically greater than with classic 4.6 mm columns.

The Van Deemter equation: peak broadening within the column

The Van Deemter equation describes how the flow velocity of the mobile phase affects the efficiency of the column. The equation reads: H = A + B/u + Cu, where H is the plate height (the smaller, the more efficient) and u is the linear flow velocity.

TermNameWhat it describesDominant at
AEddy diffusionMolecules follow different flow paths through the packed bedAll flow velocities; depends on packing quality
B/uLongitudinal diffusionMolecules diffuse from high to low concentrationLow flow velocities
CuMass transferAnalytes need time to reach equilibrium between phasesHigh flow velocities

Mnemonic

A stands for Alternative paths. B for Brownian diffusion (longitudinal diffusion). C for Contact between mobile and stationary phase (mass transfer). Remember: too slow gives diffusion, too fast gives poor mass transfer. The optimum lies in between.

In practical terms, this means that for every column there is an optimal flow velocity at which the plate height is minimal and the separation maximal. Smaller particles shift this optimum to higher flow velocities and flatten the curve, which explains why UHPLC still works efficiently at higher flows.

Recognising and diagnosing peak broadening

Before taking measures, it is important to establish where the broadening comes from. This table helps with the diagnosis.

ObservationProbable causeFirst action
All peaks broad, including early peaksExtra-column dead volumeCheck fittings, shorten capillaries, smaller detector cell
Only late peaks broadNormal diffusion effect or flow too lowOptimise the gradient or increase the flow towards the optimum
Peaks broad and tailingDead volume or secondary interactions on the columnCheck the column inlet connection; consider adjusting the pH
Sudden broadening after maintenanceIncorrectly mounted fitting or ferruleDisassemble and check that the ferrule is at the right depth
Gradual broadening over weeksColumn degradation or blocked fritFlush the column; replace if there is no recovery
Broadening with a narrow column (2.1 mm)System not optimised for low volumesMinimise the internal volume of the entire flow path

Seven measures to reduce peak broadening

  1. Use capillaries with the smallest possible internal diameter between injector, column and detector. The difference between 0.25 mm and 0.12 mm ID is immediately visible in the peak shape with UHPLC.
  2. Keep capillaries as short as practically possible. Every excess centimetre adds volume.
  3. At every fitting, check that the capillary is fully inserted up to the sealing surface. A gap of half a millimetre creates a measurable dead volume.
  4. Choose a detector cell with a volume that matches your column diameter. A large cell on a narrow column destroys the gain in column efficiency.
  5. Dissolve your sample in the mobile phase, not in a stronger eluent. Injecting in a stronger solvent causes band broadening at the column head.
  6. Limit the injection volume. Too large an injection volume overloads the column head and broadens the starting band.
  7. Work around the optimal flow velocity of your column according to Van Deemter, not at a standard 1.0 ml/min.

Never use Teflon tape on HPLC connections

Teflon tape or other sealants on HPLC fittings create dead volume and introduce particles into the system. HPLC fittings seal via the sealing surface or the ferrule, not via the thread.

Frequently asked questions about peak broadening

What is the difference between peak broadening and tailing?

Peak broadening makes the peak symmetrically broader: the peak retains its Gaussian shape but is less sharp. Tailing is an asymmetrical distortion in which the back of the peak trails off. Tailing more often indicates secondary interactions with the stationary phase or dead volume at the column inlet, while general broadening is usually system-wide.

How much dead volume is acceptable?

That depends on your column diameter. Rule of thumb: the total extra-column volume may be no more than 10 percent of the peak volume of your narrowest peak. With a 4.6 mm column, a few tens of microlitres is often acceptable; with a 2.1 mm UHPLC column, you must stay below five microlitres.

Why do my peaks become broader at lower flow?

At low flow velocities, the B term from the Van Deemter equation dominates: longitudinal diffusion. Molecules simply get more time to diffuse out into the mobile phase. Increase the flow towards the optimum of your column to limit this effect.

Our partners
  • imgi_2_Ashtech-500x73.png
  • imgi_3_asi-500x308.png
  • imgi_4_Biomiga-500x155.png
  • imgi_5_biotech-500x308.png
  • imgi_6_brown-500x308.png
  • imgi_1_logo_01.png
  • imgi_7_crs-500x308.png
  • imgi_8_Diba-logo-500x500.jpg
  • imgi_9_hamilton-500x308.png
  • imgi_10_harvard-500x308.png
  • imgi_12_idex.jpg
  • imgi_13_ismatec-500x308.png
  • imgi_14_kdscientific-500x308.png
  • imgi_15_Lablids-foto-500x133.png
  • imgi_16_newera-500x308.png
  • imgi_17_oakwade-500x308.png
  • imgi_18_Omnifit-logo-500x187.png
  • imgi_19_optimizetechnologies-500x308.png
  • imgi_20_polylc-500x308.png
  • imgi_21_polymicrotechnologies-500x308.png
  • imgi_22_quadrex-500x308.png
  • imgi_23_sciencix-500x308.png
  • imgi_24_vaplock-500x308.png
  • imgi_25_vici-500x308.png
  • imgi_26_biochem-500x308.png
  • imgi_27_chemtec-500x308.png
  • imgi_28_Drastar-logo.png
  • imgi_29_Eisele-logo-500x169.png
  • imgi_30_goregular-500x308.png
  • imgi_31_humphrey-500x500.jpg
  • imgi_32_Key-Instruments.png
  • imgi_33_KOFLOC_E_02_blue_L-442x500.png
  • imgi_34_micrafilter-500x308.png
  • imgi_35_parker-500x308.png
  • imgi_36_logo-psg.png
  • imgi_37_permapure-500x308.png
  • imgi_38_Logo-Salaera-500x278.jpg
  • imgi_39_Vogtlin-logo.png
See all partners
Receive our newsletter!

Subscribe to our newsletter to receive our latest news

Continue shopping Quotationlist
empty_cart

Your quotation list is still empty!

Add products to your quotationlist to request a quotation
  • :
  • :